How can I interpret a standard 12‑lead ECG recorded at 25 mm/s and 10 mm/mV in an adult?

Medical Advisory BoardAll articles are reviewed for accuracy by our Medical Advisory Board
Educational purpose only • Exercise caution as content is pending human review
Article Review Status
Submitted
Under Review
Approved

Last updated: February 17, 2026View editorial policy

Personalize

Help us tailor your experience

Which best describes you? Your choice helps us use language that's most understandable for you.

How to Read a Standard 12-Lead ECG

Master ECG interpretation by following a systematic approach that begins with verifying proper calibration (25 mm/s paper speed and 10 mm/mV gain), then sequentially analyzing rate, rhythm, axis, intervals, and morphology across all 12 leads. 1

Step 1: Verify Calibration and Technical Quality

Before interpreting any ECG, you must confirm standardization:

  • Check the calibration pulse – it must produce exactly a 10 mm vertical deflection, confirming the standard 10 mm/mV gain. 1 If this pulse is incorrect, all voltage measurements become unreliable and diagnostic criteria for conditions like left ventricular hypertrophy or ST-segment elevation will be invalid. 1

  • Confirm paper speed is 25 mm/s – at this speed, each small 1 mm box represents 0.04 seconds and each large 5 mm box represents 0.2 seconds, enabling precise interval measurements. 1

  • Verify temporal alignment – digital ECGs should maintain alignment of all leads within 10 ms maximum misalignment to preserve accurate spatial-temporal analysis. 2

Step 2: Understand the ECG Layout

The standard format presents four columns of 2.5-second recordings:

  • Column 1: Limb leads I, II, III (simultaneous)
  • Column 2: Augmented leads aVR, aVL, aVF (simultaneous)
  • Column 3: Precordial leads V1, V2, V3 (simultaneous)
  • Column 4: Precordial leads V4, V5, V6 (simultaneous)
  • Additional rhythm strip: Usually a continuous 10-second recording of one lead (commonly lead II) for rhythm analysis 2

The simultaneous acquisition allows precise temporal alignment of waveforms from different leads, which provides critical diagnostic value for identifying fascicular blocks, arrhythmias, and myocardial infarction. 2

Step 3: Calculate Heart Rate

Using the 25 mm/s paper speed:

  • Regular rhythms: Count the number of large boxes (5 mm) between consecutive R waves and divide 300 by this number (e.g., 3 boxes = 300÷3 = 100 bpm). 1

  • Irregular rhythms: Count the number of QRS complexes in a 6-second strip (30 large boxes) and multiply by 10. 1

Step 4: Assess Rhythm

  • Identify P waves – look for consistent P wave morphology before each QRS complex in lead II. 3

  • Measure P-P intervals – determine if the atrial rhythm is regular or irregular. 3

  • Measure R-R intervals – determine if the ventricular rhythm is regular or irregular. 3

  • Assess P-QRS relationship – confirm each P wave is followed by a QRS complex with a consistent PR interval. 3

Step 5: Determine Axis

The QRS axis represents the overall direction of ventricular depolarization:

  • Normal axis: -30° to +90° (positive QRS in leads I and aVF). 3

  • Left axis deviation: -30° to -90° (positive in lead I, negative in aVF) – suggests left anterior fascicular block or left ventricular hypertrophy. 2

  • Right axis deviation: +90° to +180° (negative in lead I, positive in aVF) – suggests right ventricular hypertrophy or left posterior fascicular block. 3

The temporal alignment of waveforms in aVR and aVL aids in diagnosing fascicular blocks, particularly in the presence of infarction. 2

Step 6: Measure Intervals

Using the 0.04-second small boxes:

  • PR interval (start of P to start of QRS): Normal is 0.12–0.20 seconds (3–5 small boxes). Prolongation indicates first-degree AV block. 3

  • QRS duration (start to end of QRS complex): Normal is <0.12 seconds (<3 small boxes). Widening suggests bundle branch block or ventricular conduction delay. 4

  • QT interval (start of QRS to end of T wave): Must be corrected for heart rate (QTc). Prolongation increases risk of torsades de pointes. 3

Step 7: Analyze Waveform Morphology

P Waves

  • Normal: <0.12 seconds wide, <2.5 mm tall in limb leads. 3
  • Abnormalities: Tall peaked P waves suggest right atrial enlargement; wide notched P waves suggest left atrial enlargement. 4

QRS Complex

  • Q waves: Pathological Q waves (≥0.04 seconds wide or ≥25% of R wave height) indicate prior myocardial infarction. 4
  • R wave progression: R waves should progressively increase from V1 to V6; poor progression suggests anterior infarction. 3
  • Bundle branch blocks: Right bundle branch block shows RSR' pattern in V1; left bundle branch block shows broad monophasic R waves in V5-V6. 4

ST Segments

This is where proper calibration is absolutely critical:

  • ST elevation: At the 10 mm/mV standard gain, ≥2 mm elevation in V1-V3 (for men ≥40 years) or ≥1 mm in other leads indicates acute myocardial infarction requiring immediate intervention. 1

  • ST depression: ≥1 mm (0.1 mV) horizontal or downsloping depression suggests myocardial ischemia. 1

Without consistent 10 mm/mV calibration, these voltage-based diagnostic criteria become unreliable and potentially dangerous. 1

T Waves

  • Normal: Upright in leads I, II, V3-V6; inverted in aVR. 3
  • Abnormalities: Deep symmetric T wave inversions suggest ischemia; peaked T waves suggest hyperkalemia. 3

Step 8: Synthesize Findings Across All 12 Leads

The 12 leads provide different anatomical views:

  • Inferior wall (II, III, aVF): Right coronary artery territory. 5
  • Lateral wall (I, aVL, V5-V6): Left circumflex artery territory. 5
  • Anterior wall (V1-V4): Left anterior descending artery territory. 5
  • Septal (V1-V2): Proximal left anterior descending. 5

Simultaneous views of P waves and QRS waveforms in multiple leads add critical information for interpreting arrhythmias and diagnosing myocardial infarction. 2

Common Pitfalls to Avoid

  • Never use torso-placed limb electrodes for diagnostic ECGs – Mason-Likar and other monitoring lead placements alter QRS morphology and can produce false-positive or false-negative infarction criteria. These tracings cannot be used interchangeably with standard ECGs for serial comparison. 2

  • Beware of filtering artifacts – reducing the high-frequency cutoff to 40 Hz (to reduce noise) will invalidate amplitude measurements used for diagnostic classification. The high-frequency cutoff must be at least 150 Hz for adults and adolescents. 2

  • Check for motion artifact – noise from limb movement can mimic arrhythmias or ST-segment changes. 2

  • Verify electrode placement – misplaced precordial leads (especially V1-V2) can simulate bundle branch blocks or infarction patterns. 5

Clinical Application

The standardized 25 mm/s and 10 mm/mV settings ensure consistent interpretation of diagnostic criteria across different ECG machines and institutions, facilitating reliable patient care. 1 Digital electrocardiographs that analyze averaged beats show significantly less variability than those measuring single complexes, improving measurement reproducibility. 2

For wide-QRS-complex tachycardias, the standard electrode sites and preserved QRS-axis criteria enable accurate distinction between ventricular tachycardia and supraventricular tachycardia with aberrant conduction. 4

References

Guideline

ECG Calibration Standards for Accurate Diagnostic Interpretation

Praxis Medical Insights: Practical Summaries of Clinical Guidelines, 2026

Guideline

Guideline Directed Topic Overview

Dr.Oracle Medical Advisory Board & Editors, 2025

Research

Basics of the 12-lead ECG.

Nursing, 2023

Research

An introduction to electrocardiogram interpretation: part 2.

Nursing standard (Royal College of Nursing (Great Britain) : 1987), 2009

Professional Medical Disclaimer

This information is intended for healthcare professionals. Any medical decision-making should rely on clinical judgment and independently verified information. The content provided herein does not replace professional discretion and should be considered supplementary to established clinical guidelines. Healthcare providers should verify all information against primary literature and current practice standards before application in patient care. Dr.Oracle assumes no liability for clinical decisions based on this content.