ECG Calibration Standards
Before each ECG recording, the device must display a standard calibration pulse (square-wave deflection) at 10 mm/mV amplitude and 25 mm/second paper speed, with these settings clearly marked on every tracing to ensure accurate voltage and timing measurements. 1
Standard Calibration Requirements
Voltage Calibration
- Set the standard gain to 10 mm/mV, where 1 mV produces a 10 mm deflection on the ECG paper 1, 2
- Each small box (1 mm) represents 0.1 mV and each large box (5 mm) represents 0.5 mV 1
- The calibration pulse must be visible at the beginning or end of each tracing to confirm proper recorder function before interpretation 1
- Voltage calibration must be accurate to ±2% to ensure reliable amplitude measurements 3
Paper Speed and Timing
- Standard paper speed is 25 mm/second, allowing accurate measurement of intervals critical for detecting conduction abnormalities and infarction patterns 2
- At this speed, each small box (1 mm) represents 0.04 seconds and each large box (5 mm) represents 0.20 seconds 1
Documentation Requirements
- Every printed or digitally saved ECG strip must show voltage (mm/mV) and paper-speed (mm/s) markings to ensure traceability 1
- When ECGs are stored or transferred within electronic health records, voltage standardization must remain visible to avoid misinterpretation of voltage-dependent criteria 1
- Always document if non-standard calibration (5 mm/mV or 20 mm/mV) is used, as this affects voltage-based diagnostic criteria 1
Technical Specifications for Accurate Measurements
Digital Sampling Requirements
- Minimum digital sampling rate of 500 samples per second is required to support accurate 150 Hz high-frequency filtering and keep amplitude-measurement error below 1% 1
- Data should be sampled at no less than 1,000 Hz and A/D converted with at least 12-bit precision for high-resolution applications 3
- All ECG leads must be recorded and converted concurrently 3
Frequency Response Standards
- For adults and adolescents: high-frequency cutoff must be ≥150 Hz to preserve waveform fidelity and prevent systematic underestimation of QRS amplitude 3, 1
- For infants and children: high-frequency cutoff must be 250 Hz to capture rapid deflections accurately 3, 1
- Low-frequency cutoff should be 0.05 Hz for routine filters, or may be relaxed to 0.67 Hz when using linear digital filters with zero-phase distortion 1
- The minimum band pass should be from 0.5 Hz to 250 Hz 3
Amplifier Requirements
- ECG signals should be recorded with a low-noise amplifier meeting American Heart Association standards for leakage current 3
- The range of linearity for input signals should not be less than ±2.5 mV 3
- Notch filters for power line interference should not be used, as they distort waveforms 3
Electrode Preparation and Placement
Skin Preparation
- Thoroughly cleanse the subject's skin with alcohol or another solvent and abrade to decrease impedance 3
- Ideally, impedance should be measured and be less than 1,000 Ω 3
- Silver-silver chloride electrodes have the lowest half-cell potential and are the electrodes of choice 3
Lead Placement Accuracy
- Precordial lead placement variability is a major source of measurement error—variation of as little as 2 cm can result in important diagnostic errors 3
- Fewer than two-thirds of routinely applied precordial electrodes are placed within 1.25 inch of the designated landmark 3
- Serial tracings in acute or subacute care settings should use skin marking to promote reproducibility of lead placement when electrodes cannot remain in place 3
- Periodic retraining in proper lead positioning should be routine for all personnel responsible for recording ECGs 3
Critical Pitfalls and Quality Assurance
Invalidating Technical Errors
- Employing a suboptimal high-frequency cutoff (e.g., 40 Hz) invalidates all amplitude-based diagnostic measurements, such as voltage criteria for left ventricular hypertrophy 1
- Inadequate high-frequency response systematically underestimates QRS amplitude and smooths critical features like Q waves, potentially leading to missed diagnoses of myocardial infarction 1, 2
- Poor frequency response causes smoothing of pathological Q waves and notched QRS components, leading to inaccurate diagnoses 2
Lead Misplacement Detection
- Medical personnel responsible for ECG recording should receive training on avoidance and recognition of lead switches 3
- Lead-switch detection algorithms should be incorporated into digital electrocardiographs with alarms for abnormally high lead impedance 3
- Suspected misplacements should be identified to the person recording the ECG in time to correct the problem 3
- Transposition of lead wires to V1 and V2 can cause reversal of R-wave progression that simulates anteroseptal wall infarction 3
Measurement Accuracy Standards
- Digital electrocardiographs must provide beat alignment that allows selective averaging or formation of a representative complex with adequate fidelity 3
- Noise levels should be reduced to <1 µV (root mean square) in the averaged signal for time-domain analysis 1
- Trigger jitter should be less than 1.0 msec and ideally 0.5 msec to prevent attenuation of high-frequency signals during signal averaging 3
Automated Alerts
- ECG devices should automatically generate alerts when suboptimal filter settings are detected and restore appropriate settings before the next recording 1
- A diagnostic statement alerting the reader to the presence of different types of lead switches should be incorporated into preliminary interpretive reports if not corrected before recording 3
Age-Specific Calibration Protocols
Pediatric Recordings
- Pediatric ECG acquisition must employ standard 10 mm/mV gain plus a 250 Hz high-frequency cutoff to avoid loss of rapid components in the QRS complex 1
- The higher frequency response is essential because children have more rapid deflections that would be smoothed by adult filter settings 3, 1
Adult Recordings
- For adults >35 years, use standard 10 mm/mV calibration with 150 Hz high-frequency cutoff 1
- For adolescents (16-35 years), use standard calibration but recognize that voltage criteria for left ventricular hypertrophy have lower accuracy in this age group 1
Serial Comparison Requirements
- Maintain consistent calibration across all ECGs to track evolution of ischemic changes or development of new infarction 2
- Compare current ECG with baseline using identical calibration settings to detect interval changes 2
- Use the largest-amplitude deflection in each lead as representative of that measurement, per ANSI/AAMI standards 1